| Pressure Measurement Specifications | 
| Operating Pressure Range | -14.7 to +300 psi Option available for absolute pressures via Barometric Reference sensor. See Option A
 | 
| Pressure Scale Units | PSI, ft. H2O, in Hg, kPa standard - 23 different scales under SET-UP, user selectable | 
| Overpressure | To 1.5x causing negligible calibration change | 
| Pressure Media | Compatible with most common non-corrosive gasses | 
| Transduction Principle | Dual, integrated silicon strain gauge sensors. Optional Barometric Reference utilizes vibrating silicon element. | 
| Readout | ¡¾999999 capability on both pressure and electrical parameters. ¡¾9999 for ranges below 1.5 psi (FS) | 
| Display | 1.5" x 5.2" L.C.D. graphics panel. 240 x 64 pixels. Backlit. | 
| Display Overload | 70% F.S. Pressure overrange nominal. Above this the display flashes. | 
| Resolution | 17 bits (0.00075% F.S.) | 
| Response | 2 readings per second nominal | 
| Zero Control | Pressure offset correction by key press Ability under process key to permit TARE of pressure, voltage and current, either manual or displayed value. | 
| Accuracy | ¡¾0.025%, of reading between 1% and 100% F.S. 0.002 psi below 1% F.S. Figure includes 90 day stability. Combined non linearity, hysteresis and repeatability (included within the above accuracy figure). | 
| Positive Pressures | ¡¾0.025% of reading between 1% and 100% F.S. ¡¾0.001 psi below 1% F.S. | 
| Negative Pressures | ¡¾0.025% of reading between -14.7 psi to -3 psi ¡¾0.001 psi between 0 and -3 psi | 
| Temperature Effects | -14.7 to 300 psi gauge calibrator over 15¡Æ to 100¡ÆF the averaged temperature coefficient ¡¾0.0012% of reading/¡ÆF | 
| Position Effect | Negligible | 
| Electrical Specifications | 
| Electrical Inputs | 
| Voltage Measurement | 0 to ¡¾50 V DC auto ranging, resolution 10¥ìV. Input impedance 10m¥Ø. Accuracy ¡¾0.04% of reading ¡¾1 digit including 90 day stability. ¡¾0.004% of reading/¡ÆF | 
| Power Supply | Supplied fitted with Ni-Cad rechargeable pack. Can also be powered by 6 x D cells, alkaline recommended. Battery Life: 25 hours nominal for Ni-Cad. 75 hours nominal for alkaline. External PSU/charger unit suppliedas standard. | 
| Current Measurement | 0 to ¡¾55 mA d.c. resolution 0.001mA Internal resistance nominally 10¥Ø/ Accuracy 0.03% of reading ¡¾ 1 digit including 90 day stability. ¡¾0.004% of reading/¡ÆF | 
| Electrical Outputs | 
| Voltage Output | Programmable floating output in the range 0 to 24 V DC max 50 mA. Suitable also for current loop energizing and switch testing. | 
| Current Output | Programmable output in the range 0 to 55 mA DC for current display calibrations. Configured as source, voltage limit 19 V. Configured as sink, voltage limit maximum 30 V, voltage limit minimum 5V. | 
| Analog Output Signal | Proportional to displayed pressure. Accuracy s display. 0 to 24 V DC with 16 bit (0.001% F.S.) resolution. User programmable zero and full scale settings. Bandwidth 1 to 2 Hz - updated each new reading. Current output can also be provided either 0 to 50mA or 4 to 20mA for 0 to F.S. pressure | 
| Environmental Specifications | 
| Temperature | Operating 15¡Æ to 100¡ÆF Storage 0¡Æ to 140¡ÆF | 
| Sealing | Instrument NEMA 4X Battery charger to laboratory standard. | 
| Electro Magnetic Compatibility | Designed to meet BS6667/HD481 and BS6527 / EN55022. | 
| Physical Specifications | 
| Dimensions/Weight | 5.5 inches z 11 inches x 14.5 inches / 9 lbs. nominal | 
| Pressure Connections | 1/8 NPT female | 
| Included | Carrying case and a complete set of test leads are included. |