These new units beat every other comparable dry-well in the industry in performance, size, weight, convenience, ease of calibration, software and price. In addition, the heating and cooling rate of each of these dry-wells is adjustable from the front panel, thermal switches can be checked by these units for actuation testing and multiple-hole inserts are available for a variety of probe sizes.

 

Choose one of three removeable inserts sized from 1/16" to 1/2" diameter. Insert A handles a full range of probe sizes with a single well of each size. Insert B features two wells each of 3/8", 1/4" and 3/16" in diameter for doing comparison calibrations, and insert C has six 1/4" diameter wells for multiple probe calibrations
 
Features
  • Accuracy to ±0.5°C
  • RS-232 and Windows® software included
  • Easy to recalibrate
  • Less than eight pounds
 
SPECIFICATIONS — Series ZB

Range

-25 to 140°C (-13 to 284°F), in ambient of 23°C
±3°C (Can do lower depending on ambient.)
Accuracy ±0.25°C
Stability ±0.02°C @ -25°C, ±0.04°C @ 140°C
Uniformity ±0.3°C typical (±0.1 with similar size wells)
Resolution 0.1°
Display Scale °F or °C, switchable
Heating Time 18 minutes, ambient to 140°C
Cooling Time 20 minutes, ambient to -25°C
Stabilization Time 7 minutes
Inserts (included) Insert A: 1/2", 3/8", 1/4", 3/16", 1/8" and 1/16" holes
Depth 4.875 inches (124 mm) immersion
Computer Interface RS-232 included
Size 10.25" H x 5.63" W x 9.63" D (26.1 x 14.3 x 24.5 cm)
Weight 12.5 lb. (5.7 kg)
Certification 7-point NIST-traceable certificate of calibration included

SPECIFICATIONS

Model

Series ZA (shown right) Series ZC (shown left)
Range 95°F to 662°F (35°C to 350°C) 122°F to 1202°F (50°C to 650°C)
Accuracy ±0.9°F (±0.5°C); accuracy in insert holes greater than 1/4" will be at ±1.8°F (±1°C)  122°F to 752°F (50°C to 400°C):
±0.9°F (±0.5°C), 752°F to 1202°F (400°C to 650°C):
±1.8°F (±1°C), holes greater than 1/4" (6.35 mm): ±3.6°F (±2°C)
Stability ±0.03°C @ 50°C; ±0.05°C @ 100°C; ±0.10°C @ 500°C;
±0.05°C @ 350°C ±0.12°C @ 650°C
Uniformity ±0.72°F (±0.4°C) typical ±0.18°F (±0.1°C) with similarly sized wells 122°F to 752°F (50°C to 400°C): ±0.72°F (±0.4°C) typical, ±0.9°F (±0.5°C) with similarly sized wells above 752°F (400°C)
Heating Times 12 minutes from ambient to 662°F (350°C) 12 minutes from ambient to 1202°F (650°C)
Cooling Times 15 minutes from 662°F to 212°F (350°C to 100°C) 25 minutes from 1202°F to 212°F (650°C to 100°C)
Stabilization Time 7 minutes
Immersion Depth 4.875 inches (124 mm)
Computer Interface RS-232 included with free Windows®-based software (Model 9930)
Power 115 VAC (±10%), 50/60 Hz and 230 VAC (±10%), 50/60 Hz switchable
Size 6" W x 3.375" H X 7.75" D (152 x 86 x 197 mm) 4.5" W x 9" H x 7" D (114 x 229 x 178 mm)
Weight 6 lb. (2.7 kg) 8 lb. (3.6 kg)
Certification Includes 7-point, NIST-traceable certificate of calibration with data at the following points: 50°C, 100°C, 150°C, 200°C, 250°C, 300°C and 350°C Includes 7-point, NIST-traceable certificate of calibration with data at the following points:100°C, 200°C, 300°C, 400°C, 500°C and 600°C

CAT NO. MODEL DEscRIPTION AVAIL PRICE 
HR9140A   9140-A   9140 FIELD DRY WELL CALIBRATOR   Yes   4,525,740
HR9140B   9140-B   DRY-WELL (B INSERT INCLUDED)   Yes   4,525,740
HR9140C   9140-C   DRY-WELL (C INSERT INCLUDED)   Yes   4,525,740
HR9141A   9141-A   9141 FIELD DRY WELL CALIBRATOR   Yes   6,481,080
HR9141B   9141-B   DRY WELL (B INSERT INCLUDED)   Yes   6,481,080
HR9141C   9141-C   DRY WELL (C INSERT INCLUDED)   Yes   6,481,080
HR31401   3140-1   OPTIONAL INSERT BLANK FOR DRY WELL FOR Field Temperature Dry-Well Calibrators   Yes   302,940

 

HR31402   3140-2   OPTIONAL INSERT A FOR DRY WELL FOR Field Temperature Dry-Well Calibrators   Yes   348,840

 

HR31403   3140-3   OPTIONAL INSERT B FOR DRY WELL FOR Field Temperature Dry-Well Calibrators   Yes   348,840

 

HR31404   3140-4   OPTIONAL INSERT C FOR DRY WELL FOR Field Temperature Dry-Well Calibrators   Yes   348,840

 

HR9308   9308   OPTIONAL RUGGED CARRYING CASE FOR Field Temperature Dry-Well Calibrators   Yes   128,520

 

HR31411   3141-1   OPTIONAL INSERT-BLANK FOR DRY WELL FOR Field Temperature Dry-Well Calibrators   Yes   459,000

 

HR31412   3141-2   OPTIONAL INSERT A FOR DRY WELL FOR Field Temperature Dry-Well Calibrators   Yes   660,960

 

HR31413   3141-3   OPTIONAL INSERT B FOR DRY WELL FOR Field Temperature Dry-Well Calibrators   Yes   660,960

 

HR31414   3141-4   OPTIONAL INSERT C FOR DRY WELL FOR Field Temperature Dry-Well Calibrators   Yes   660,960
HR9309   9309   OPTIONAL RUGGED CARRYING CASE FOR Field Temperature Dry-Well Calibrators   Yes   688,500

 

번호 제목 작성자 작성일 평가점수
이 상품에 대한 사용후기가 아직 없습니다.
사용후기를 작성해 주시면 다른 분들께 많은 도움이 됩니다.
 
* 이 상품을 사용해 보셨다면 사용후기를 써 주십시오.
번호 제목 작성자 작성일 답변
이 상품에 대한 질문이 아직 없습니다.
궁금하신 사항은 이곳에 질문하여 주십시오.
 
* 이 상품에 대한 궁금한 사항이 있으신 분은 질문해 주십시오.
이 상품과 관련된 상품이 없습니다.
DAVIS 종합 계측기 - 한국대리점 코린스계기(주)
 
회사소개 | 서비스이용약관 | 개인정보취급방침
주소 : 경기도 성남시 중원구 둔촌대로 474, 708호 (상대원동 선텍시티1)
Tel : 031-777-1588 / Fax : 031-777-1587 / 개인정보관리책임자 : 관리자
사업자 등록번호 : 116-81-31753 / 통신판매업신고 : 2005-경기성남-0886호
대표자 : 홍영화 / 운영자 : 윤상미
상호 : 주식회사 코린스 / Copyright ⓒ 2002. All Rights Reserved.
가입사실확인